Home > Current Transformers/Transducers > Understanding Current Transformer Accuracy Classes

Understanding Current Transformer Accuracy Classes

IEC 61869-2 defines the new current transformer accuracy classes intended to replace the old standard, IEC60044-1 (note that IEC 61869-1 is designed for instrument transformers) .  The new and old standards are essentially identical, but IEC 61869-2 consolidated two parts of the older standard:

  • IEC 60044-1 : Instrument transformers – Part 1: Current transformers
  • IEC 60044-6 : Instrument transformers – Part 6: Requirements for protective current transformers for transient performance

While IEEE/ANSI C57.13 remains active, most current manufacturers prefer the IEC standard because it is not specific to instrument transformers with a 5A output.  Therefore, instead of dealing strictly with 5A output devices, the IEC standard is more general, covering devices with a mA output or voltage output.

The next thing to be aware of is that the 0.5 Class (of IEC 61869-2) and the 0.5s Class (of IEC 61869-2) are different.  The 0.5s version is a more rigorous standard when it comes to performance on lower current ratios.  For example, at 5% of the rated current, the percentage current ratio error for the 0.5 Class is 1.5% whereas 0.5s requires 0.75% or better.

The biggest single mistake I see people make when comparing accuracy of CTs is to assume that 0.5% accuracy = IEC 61869-5 0.5 Class (or some other standard).  THIS IS NOT NECESSARILY THE CASE!

I recently tested several CTs from a manufacturer (that I’ll leave unnamed).  The website specified the CTs were 0.5% accurate from 10% to 120% of the rated current.  I tested the CTs and I found they were correct.  However, they were not 0.5 Class!  Here’s why:

IEC 61869-2 0.5 Accuracy

This particular unit, which was rated at 20A, has an error at 20A that is below 0.5%.   In addition, I tested the part at 10%, 25%, 50%, 75% and 120% of the rated current, and in each case the linearity was below 0.5%.  Therefore, the manufacturer’s claim is true; however, the unit’s phase shift is too large to meet the 0.5 Class.  This is because it would have to be less than 30 minutes (0.5 degrees) at 100% of the rated current.

In conclusion, 0.5% accuracy ≠ 0.5 Class ≠ 0.5s Class.  Be sure to look at the standard the CT complies with when comparing products.  And, if it doesn’t state any standard, be wary!

Advertisements
  1. No comments yet.
  1. No trackbacks yet.

Leave a Reply

Fill in your details below or click an icon to log in:

WordPress.com Logo

You are commenting using your WordPress.com account. Log Out / Change )

Twitter picture

You are commenting using your Twitter account. Log Out / Change )

Facebook photo

You are commenting using your Facebook account. Log Out / Change )

Google+ photo

You are commenting using your Google+ account. Log Out / Change )

Connecting to %s

%d bloggers like this: